3. Validation

A. Equipment Validation
  • All pipettors: calibrate every 120 days.
  • Biomek Workstation: Preventive maintenance and calibration performed annually by a Biomek technician.
  • ELISA plate reader: Autocalibration will be performed before each plate reading. In addition the linearity of detection will be tested weekly using a calibration test plate (MedTec Biolab Equipment Company).
B. Assay Solutions
  • Assay buffer: Stored at 4°and used before expiration date
  • Wash buffer: Stored at 4°and used before expiration date
  • Streptavidin polymer-peroxidase conjugate buffer: Purchased complete, stored at 4° and used before expiration date.
  • Lot numbers, date made, initials of technician, name of solution is recorded in both the solutions manual and on the solution bottle.

C. Performance of each assay:
The following information is calculated, provided to the customer with the results of each assay and recorded in our laboratory assay performance log book:

  • A mid-level internal control is included in each assay (12.5 to 25% of the highest standard concentration).
  • The inter-test %CV for the control is calculated based on an up-to-date mean of the measured internal control values.
  • An intra-test %CV is calculated for each dilution of standard, each sample, and the internal control.
  • Standard curve equation and regression coefficient (s).
  • Percent error of internal control based on an up-to-date mean of the measured internal control values; if the error exceeds 20%, the test is rejected and the assay is repeated.
  • Each assay is tested at least every 60 days to insure adequate performance.